Technical data – KROHNE OPTIGAS 4010 C EN User Manual
Page 19
TECHNICAL DATA
6
19
OPTIGAS 4010C
www.krohne.com
07/2014 - 4002962702 - MA OPTIGAS R02 en
ATEX (acc. 94/9/EC) temperature limits
Interface cable
2 x screened twisted pairs with a minimum 20 AWG conductors. Total C ≤
50nF, total L ≤ 200µH
Cable glands
M20 X 1.5 Stainless Steel suitable for cable diameter 6.5...9.5 mm
Approvals
Custody transfer
PTB type approval 5.411 / 04.15
Vibration
IEC 60068-2-6
ATEX (acc. 94/9/EC)
ATEX (acc. 94/9/EC)
ATEX (acc. 94/9/EC)
ATEX (acc. 94/9/EC)
OPTIGAS 4010C without heating jacket / insulation
OPTIGAS 4010C without heating jacket / insulation
OPTIGAS 4010C without heating jacket / insulation
OPTIGAS 4010C without heating jacket / insulation
II 1/2 G Ex ib IIC T4...T1 Ga/Gb
II 2 D Ex ib IIIC T210°C Db
Ex ib IIC T4...T1 Ga/Gb
Ex ib IIIC T210°C Db
IECEx
IECEx
IECEx
IECEx
Ex ib IIC T6...T1 Ga/Gb
Ex ib IIIC T4*** °C Db
NEPSI
NEPSI
NEPSI
NEPSI
OPTIGAS 4010C
OPTIGAS 4010C
OPTIGAS 4010C
OPTIGAS 4010C
Ex ib IIC T4...T1 Ga/Gb
1 Toolbox software is available from the manufacturer but it requires a compatible Modbus to PC adapter with approved barrier. Please
call for more information.
Ambient temp.
Tamb °C
Max. medium
temp. Tm °C
Temp. class
Max. surface
temp. °C
OPTIGAS 4010C without heating jacket /
insulation
-40...+65
60
T4
T130
125
T3
T195
140
T2 - T21
T210
MA_OPTIGAS_4010C_en_140722_4002962702_R02.book Page 19 Tuesday, July 22, 2014 4:30 PM