Technical data, 4 measuring ranges – KROHNE GA 24 EN User Manual
Page 24
6
TECHNICAL DATA
24
GA24
www.krohne.com
02/2011 - 4001229901 MA GA24 R01 en
6.4 Measuring ranges
Measuring span 10 : 1
Flow values 100%
Reference condition:
Water 20°C / Air 20°C - 1.013 bar abs.
Float materials:
1 Stainless Steel or Hastelloy® - 2 PTFE - 3 TFM - 4 Aluminium - 5 Polypropylen (PP)
Materials →
1
2
3
1
3
4
5
1
2
3
4
5
Cone ↓
Flow water [l/h]
Flow air [m
3
/h]
Max. pressure loss [mbar]
N 18.07
DN15
40
25
13
1.5
0.6
0.8
0.5
9
6
2
3
1
N 18.09
63
40
22
2.2 0.95
1.2
0.7
9
7
3
3
2
N 18.13
100
63
35
3
1.5
1.8
1.2
9
8
3
4
2
N 19.09
160
100
55
5
2.2
2.8
1.8
13
9
4
5
2
N 19.13
250
160
85
8
3.3
4.5
2.8
16
11
4
5
2
N 19.19
400
250
140
-
-
-
-
21
14
5
-
-
N 19.26
630
400
230
-
-
-
-
27
17
6
-
-
N 21.09
DN25
630
400
230
18
1
9
11
7
22
14
6
8
3
N 21.13
1000
630
350
28
1
14
18
12
23
17
6
8
4
N 21.18
1600
1000
600
49
1
-
28
1
17
1
26
25
7
10
6
N 21.25
2500
1600
950
70
1
-
42
1
26
1
33
40
8
12
9
N 41.09
DN40
1600
1000
600
45
1
22
28
18
32
18
9
11
5
N 41.13
2500
1600
900
70
1
36
45
1
28
1
34
20
10
12
5
N 41.19
4000
2500
1500
128
1
-
76
1
46
1
38
24
11
15
8
N 51.10
DN50
4000
2500
1500
120
1
56
70
45
43
25
12
15
7
N 51.15
6300
4000
2400
190
1
90
110
1
70
1
47
30
13
16
7
N 51.21
10000
6300
3500
310
1
-
170
1
118
1
55
42
14
20
10
1 only possible with guided float
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate.
Conversion to other products or operating data is performed by KROHNE using the calculation
method stipulated by VDI/VDE Directive 3513.
MA_GA24_R01_en_4001229901_PRT.book Page 24 Wednesday, February 2, 2011 8:57 AM