beautypg.com

KROHNE OPTISENS CAC 050 EN User Manual

Page 9

background image

Manual OPTISENS CAC 050 / IAC 050

9

OPTISENS CAC 050 / IAC 050

2.1

Connection diagram panel-mounting converter




Connection

Termin
als

Notes

2 = Inner electrode = white
3 = Outer electrode = brown

Conductivity sensor
conductive (CAC)

2 + 3

5 + 6

Integrated Pt 100 = yellow & green

Conductivity sensor
inductive (IAC)

1 – 4



5 + 6

Measurement = cable #3
1 = screen, 2 = core cable #3
Voltage supply = cable #1
3 = screen, 4 = core cable #1
Integrated NTC = cable #2
5 = screen, 6 = core cable #2

Pt 100 (CAC) / NTC (IAC)

5 + 6

Analog output

11 + 12

9 = + , 10 = - , max. burden 500 Ohm

Relay 1

18 + 19

Relay 2

20 + 21

Relay 3

22 + 23

Alarm relay

Power supply

24 – 26

check information given on name plate

RS485 (option)

Sub-D

3 = +, 8 = -
4/7 bridged activates terminating resistance

NOTE

When using different sensors or cables refer to the instructions for connection and cable color specifications