beautypg.com

KROHNE OPTISENS CAC 050 EN User Manual

Page 13

background image

Manual OPTISENS CAC 050 / IAC 050

13

OPTISENS CAC 050 / IAC 050

3.3

Password and language

Main menu

.

.

Enter password

Enter password

056 Code

3.3.1

Enter password

To get access to the various parameters you have to enter the correct password:

• Code 11 gives access to the parameters „calibration“, „temperature compensation“, and „set points“.

• Code 86 gives access to all parameters and functions.
With any other number it is impossible to select, view or change any parameter.

Main menu

.

.

Basic settings

Basic settings

.

.

Language

Language

We speak
English

3.3.2

Language

For the communication with the device you can choose from a variety of languages.

Since choosing a language is part of the basic settings, it requires code 86. If a different code is set, you will
be asked to enter the correct password.