beautypg.com

KROHNE OPTISENS CAC 050 EN User Manual

Page 12

background image

Manual OPTISENS CAC 050 / IAC 050

OPTISENS CAC 050 / IAC 050

12


Enter password

058


1) Address the parameter with key

.

Enter password

058

2) A double triangle appears behind the number indicating that the
number can be changed now with keys

and

.

Enter password

062

3) Store the new value with key „Enter“. The double arrow disappears -
the new value is stored.

3.1.2

Adjustment of numerical parameters


Numerical parameters can only be altered when a double arrow is visible behind the number. This double
arrow appears when you address the parameter with key

.


Adjust the parameter with keys

and

. A short pressure on the key changes the last decimal by 1. If you

keep the key pressed, the value will continue changing until the pressure is released.

Store the changes with key „Enter“. The double arrow disappears.

3.2

Menu Overview


6.98 mS 25.0°C

S1

S2 Auto

Display of the measured values

Press key

to enter the main menu.

Main menu

Temp. comp.

Temperature compensation

Enter password

Password function

Set points

Controller settings - set points, P ranges, I functions

Limit values

Alarm function

Basic settings

Basic settings

Cal. Pt 100 / NTC

Service

Contr. settings

Turn-on delay

Analog output

Language

Temp. coeff.

Cell factor

All parameters which are set just once
at the beginning can be found in the
menu “basic settings“

Cable Comp.

Bus address

Meas. range

Averaging

3.2.1

Main menu and basic settings


The parameters are sorted into two menus: In the main menu you will find all functions which are used
regularly. The menu basic settings contain all parameters which are set just once during start-up.

On the following pages you will find information on how to adjust parameters and which parameters you
need for which application. The information follows the following scheme:
1) General adjustments: password and language
2) Adjustments for measurement: cable compensation and temperature compensation, averaging and
adjustment of the cell factor
3) Adjustments of the controller: selection of the controller version and corresponding parameters
4) Adjustments to read out data: analog, digitally and/or as alarm