Manual OPTISENS CAC 050 / IAC 050

OPTISENS CAC 050 / IAC 050

24


6.1

Error messages

Error message

Cause

Action

Error input 1

The conductivity sensor gives no
signal.

Please check the connections, the cable
and the sensor for signs of damage.

Error input 2

The temperature sensor gives no
signal.

Please check the connections, the cable
and the sensor for signs of damage. This
message also appears if automatic
temperature compensation was selected
although no temperature sensor was
used.

Limit 1 / 2

The measured value exceeded limit
1 (or dropped below limit 2,
respectively).

Please check the dosing and readjust the
control parameters, if necessary.

Dosage

Pump 1 (or 2, resp.) feeds with
100% for more than the defined
period of time.

Please check the dosing, especially the
feeding tubes and connections. Caution!
Carefully check for leaking chemicals!

Ext. controller stop

The digital input has been short-
circuited.

This only indicates the external controller
stop. If, however, you have connected a
level sensor, this message appears due
to the „lack of water“ alarm.

7.

Output

Main menu

.

.

Basic settings

Basic settings

.

.

Analog output

Selection

Analog output

Select output

-between 0-20 mA and

.

0 – 20 mA

4-20 mA

.

Begin 0/4 mA

-of the measured value

.

0.00mS

Corresp. to 4 mA

End 20 mA

-of the measured value

2.00 mS

Corresp. to 20 mA

Analog output

-between readout of the

Measuring value

Measurement or the

Control values

Serial interface

RS485 (option)

Bus address

Bus address

Nr. 00

Set a number between
0 and 31 to address the
device via data bus.