beautypg.com

KROHNE OPTISOUND 3050C 4wire EN User Manual

Page 45

background image

Electromechanical data
Cable entry
- Double chamber housing

l

1x cable entry M20x1.5 (cable-ø 5 ... 9 mm),

1x blind stopper M20x1.5

or:

l

1x closing cap ½ NPT, 1x blind stopper ½

NPT

Spring-loaded terminals

for wire cross sections up to 2.5 mm²

Transducer cable

7)

- length

5 ... 300 m (16.4 ... 984.3 ft)

- Diameter

7.2 ... 7.6 mm

Indicating and adjustment module
Power supply and data transmission

through sensor via gold-plated sliding contacts

(I²C bus)

Indication

LC display in full dot matrix

Adjustment elements

4 keys

Protection
- unassembled

IP 20

- mounted into the sensor without

cover

IP 40

Materials
- Housing

ABS

- Inspection window

Polyester foil

Supply voltage
Supply voltage
- non-Ex and Exd instrument

20 … 72 V DC, 20 … 253 V AC, 50/60 Hz

Power consumption

max. 4 VA; 2.1 W

Electrical protective measures
Protection

IP 66/IP 67

Overvoltage category

III

Protection class

I

Approvals

8)

ATEX

ATEX II 1/2D IP66 T

7)

With separate version C and D.

8)

Deviating data with Ex applications: see separate safety instructions.

OPTISOUND 3050 C - 4 ... 20 mA/HART - four-wire

45

Supplement

30518

-EN

-050914