Addendum to Operating and Instruction Manual for “Modis” Converters
KROHNE Ltd
•
Rutherford Drive
•
Park Farm Industrial Estate
•
Wellingborough
•
Northamptonshire
•
NN8 6AE
•
UK
Tel. +44 1933 408 500
•
Fax +44 1933 408 501
•
WWW.KROHNE.com
Page 13 of 18
0
1
0 0 0 0
= non-specific
0
1
0 0 0 1
= last usable value
0
1
0 0 1 0
= substitute-set
0
1
0 0 1 1
= initial value
0
1
0 1 0 0
= sensor conversion not accurate
0
1
0 1 0 1
= engineering unit violation (unit not in the
valid set)
0
1
0 1 1 0
= sub-normal
0
1
0
1
1
1
= configuration error
0
1
1
0
0
0
= simulated value
0
1
1
0
0
1
= sensor calibration
Status = good (Non-Cascade)
1
0
0 0 0 0
= ok
1
0
0 0 0 1
= update event
1
0
0 0 1 0
= active advisory alarm (priority < 8)
1
0
0 0 1 1
= active critical alarm (priority > 8)
1
0
0 1
0 0
= unacknowledged update event
1
0
0 1
0 1
= unacknowledged advisory alarm
1
0
0 1
1 0
= unacknowledged critical alarm
1
0
1
0
0
0
= initiate fail safe
1
0
1
0
0
1
= maintenance required
Status = Limits
0 0 = ok
0 1 = low limited
1 0 = high limited
1 1 = constant
Please test the first two (Quality) to check the quality of the measuring value:
Good (non Cascade)
measuring value is ok and can be used
Uncertain
the measuring value can be used, although the accuracy is not
confirmed (e. g. measuring value is frozen or A/D-value is outside
of the valid range)
Bad
the measuring value is not acceptable and cannot be used for
further processing.
Good (Cascade)
will not be supported, because it is not applicable for the device
Diagnosis
If the internal diagnostic detects a error, additional information will be transmitted in the
Master. The meaning of this information are described in the GSD-file under
UNIT_DIAG_BIT(i).