6 Limits

7. Grounding

Warning: All flowmeters must be properly grounded to avoid personnel shock hazard.

The ground conductor should not transmit any interference voltages, therefore do not ground
any other electrical devices together with this conductor.

Separate primary head IFS 6000 F with terminal box

A functional ground FE must always be connected.

Signal converters with a field power supply of > 125 mA / > 60 V:
Because of the higher field current from the signal converter, a PE protective conductor must
be connected to the IFS 6000 F primary head, see grounding diagrams.

Connections

Connection
meter size

Max. operating pressure in bar/psig at product temperature of ...

Flange DIN 2501/PN 40

Flange JIS 2210/20 K

Flange ANSI B 16.5: 150 lb

300 lb

Aseptic weld-on connection

for pipes to DIN 11 850

Aseptic weld-on connection

for pipes to ISO 2037

Dairy screw connection DIN 11 851

Screwed pipe connection ISO 2853

Screwed pipe connection SMS 1145

Clamp joint ISO 2852

DN 10*-15

DN 10*-15

1

/

2

*

1

/

2

*

DN 10*-40

DN 50-80

12*-38/1”-1

1

/

2

51-76.1/2”-3”

DN 10*-40

DN 50-80

12*-38 mm

51-76.1 mm/1

1

/

2

”-3.0”

25-76 mm

12*-51 mm/1

1

/

2

”-2.0”

63.5-76.1 mm/2

1

/

2

”-3.0”

< 40°C

(< 105°F)

39

.6

(566)

39

.6

(566)

19.6 (284)

39

.6

(566)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

0

6

.6 0

(90)

16

.6

(230)

10

.6

(150)

< 60°C

(< 140°F)

37

.6

(537)

37

.6

(537)

19.0 (276)

37

.6

(537)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

0

6

.6 0

(90)

16

.6

(230)

10

.6

(150)

< 70°C

(< 158°F)

36

.6

(522)

36

.6

(522)

18.7 (271)

36

.6

(522)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

0

6

.6 0

(90)

16

.6

(230)

10

.6

(150)

< 90°C

(< 195°F)

34

.6

(493)

34

.6

(493)

18.1 (263)

34

.6

(493)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

0

6

.6 0

(90)

16

.6

(230)

10

.6

(150)

< 100°C

(< 210°F)

33

.6

(479)

33

.6

(479)

17.7 (257)

33

.6

(479)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

0

6

.6 0

(90)

16

.6

(230)

10

.6

(150)

< 120°C

(< 250°F)

32

.6

(464)

32

.6

(464)

17.0 (247)

32

.6

(464)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

0

6

.6 0

(90)

16

.6

(230)

10

.6

(150)

< 140°C **

(< 285°F)

30

.6

(435)

30

.6

(435)

16.2 (235)

30

.6

(435)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

40

.6

(580)

25

.6

(360)

Vacuum load all sizes

0 mbar abs.

< 180°C **

(< 355°F)

28

.6

(406)

28

.6

(406)

14.7 (213)

28

.6

(406)

*

*

size DN 2.5 - 10 /

1

/

10

” -

3

/

8

** max. process temperature 140°C (285°F) for compact flowmeters

ambient temperature here max. 40°C/104°F.