beautypg.com

Description of hart interface, 8 basic-dd menu tree (details for settings) – KROHNE SMARTSENS PH 8530 EN User Manual

Page 39

background image

DESCRIPTION OF HART INTERFACE

7

39

SMARTSENS PH 8530

www.krohne.com

11/2013 - 4002544802 - MA SMARTSENS PH 8530 R02 en

7.8 Basic-DD menu tree (details for settings)

1 Measuring value

1 pH value

Display of the measured pH value on process control system, HART

®

handheld or display.

2 Temperature

Display of the measured temperature value in °C / °F on process control
system, HART

®

handheld or display.

3 Electrode voltage

Display of the measured electrode voltage in mV on process control system,
HART

®

handheld or display.

4 Loop current

Display of the measured loop current in mA on process control system,
HART

®

handheld or display.

5 Error

Display of status messages and diagnostic information. For further

information refer to

Status messages and diagnostic information on page 23.

6 Device state

Display of status information icon for sensor according to NAMUR NE 107.

2 Quick setup

1 TAG

Set the TAG number for the measuring loop.

2 Manual hold

Activate or deactivate the manual hold function. Select between yes or no.

3 I/O

Meas. value at 4 mA

Set measuring value at 4 mA;

Default setting: pH 0

Meas. value at 20 mA

Set measuring value at 20 mA;

Default setting: pH 14

Time constant

Set time constant (1…60 seconds);

Default setting: 1 second

4 Calibration

Start calibration procedure. For more information refer to

Calibration on

page 17.

5 Autoclaving counter

Rise autoclaving counter (yes/no)

6 Input user password

Set password

3 Logbooks

1 Calibration logbook

Logbook for the last 20 calibration cycles

2 Error logbook

Display errors

.book Page 39 Monday, November 18, 2013 4:32 PM