Technical data, 2 technical data – KROHNE SMARTSENS PH 8530 EN User Manual
Page 30
6
TECHNICAL DATA
30
SMARTSENS PH 8530
www.krohne.com
11/2013 - 4002544802 - MA SMARTSENS PH 8530 R02 en
6.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle
Potentiometric
Measuring range
0…14 pH
Design
Shaft diameter
12 mm / 0.47"
Insertion length
120 mm / 4.72"; 225 mm / 8.86"
Temperature sensor
Pt1000
Connector
VarioPin 2.0 (VP 2.0)
Operating conditions
Temperature range
0...+80°C / +32...+176°F
Process pressure
12 bar / 174 psi
Measuring accuracy
0.5%
Conductivity
Min. 2 µS/cm
Installation conditions
Ingress protection
IP 68
Weight
Approx. 80 g / 0.18 lb
Process connection
PG 13.5
Materials
Sensor shaft
Glass
Membrane glass
A glass
Inner buffer
pH 7.0
Reference
KCl gel
Diaphragm
3 x ceramic
Gasket
EPDM (FDA)
Sensor head
Nickel plated brass body with VP 2.0 connector
PEEK body with VP 2.0 connector (in preparation)
.book Page 30 Monday, November 18, 2013 4:32 PM