beautypg.com

Installation, 6 installing the sensor – KROHNE SMARTSENS PH 8530 EN User Manual

Page 16

background image

3

INSTALLATION

16

SMARTSENS PH 8530

www.krohne.com

11/2013 - 4002544802 - MA SMARTSENS PH 8530 R02 en

3.6 Installing the sensor

3.6.1 General installation instructions

The membrane glass must always have full contact with the measuring medium.

The mounting position of the sensor should not deviate more than 75° from vertical position
(sensor tip pointing downwards). Doing otherwise might cause internal air bubbles to float into
the membrane glass. This would interrupt the electrical contact between the inner buffer
solution and the glass surface.

Figure 3-6: Installation requirements

1 Measuring medium
2 Maximum deviation of 75° from vertical position

.book Page 16 Monday, November 18, 2013 4:32 PM