beautypg.com

Operation – KROHNE SMARTSENS PH 1590 EN User Manual

Page 21

background image

OPERATION

4

21

SMARTSENS PH 1590

www.krohne.com

11/2013 - 4002545902 - MA SMARTSENS PH 1590 R02 en

• Dip the sensor into the second pH buffer solution 3 and wait till the value is stable.
• Press Ok

Ok

Ok

Ok to proceed.

i

The measurement of pH buffer solution 2

pH buffer solution 2

pH buffer solution 2

pH buffer solution 2 starts and the values pH and temperature appear.

After approx. 10 seconds the following values appear:
pH value
temperature value
Status buffer detection → Finished
Value indentified buffer

• Press Ok

Ok

Ok

Ok to confirm the identified pH buffer solution. If neccessary change the value of the pH

buffer solution manually.

• Press Enter

Enter

Enter

Enter to proceed.

i

If the sensor does not recognise the pH buffer solution the messages
Identified buffer

Identified buffer

Identified buffer

Identified buffer → 0 pH

0 pH

0 pH

0 pH and Buffer not recognised

Buffer not recognised

Buffer not recognised

Buffer not recognised appear.

• Press Ok

Ok

Ok

Ok to proceed.

i

Enter the value of the pH buffer solution manually.

• Press Enter

Enter

Enter

Enter to proceed with the calibration procedure.

• Clean the sensor with demineralised water 2.

i

After a successful calibration the following values appear:
Buffer solution 1
Buffer solution 2
Offset in mV
Slope in mV/pH

• Press Ok

Ok

Ok

Ok to proceed.

• Set the calibration date with MM-DD-YYYY. Press the key > to enter the submenu. Now you can

change the calibration date.

• Press Enter

Enter

Enter

Enter and Ok

Ok

Ok

Ok to proceed.

i

The message save the values?

save the values?

save the values?

save the values? appears.

• Select yes

yes

yes

yes to save the values.

• Press Enter

Enter

Enter

Enter to proceed.

• Select No

No

No

No to deactivate the function manual hold

manual hold

manual hold

manual hold.

• Press Enter

Enter

Enter

Enter to leave the calibration menu.

A pH calibration is necessary in regular intervals or when installing a new pH sensor.

If the calibration was not successful, the message Warning!

Warning!

Warning!

Warning! appears behind the values Offset

Offset

Offset

Offset

and/or Slope

Slope

Slope

Slope.

• Press Ok

Ok

Ok

Ok to proceed.

• Set the calibration date with MM-DD-YYYY. Press the key > to enter the submenu. Now you can

change the calibration date.

• Press Enter

Enter

Enter

Enter and Ok

Ok

Ok

Ok to proceed.

i

The messages Warning! Offset out of range!

Warning! Offset out of range!

Warning! Offset out of range!

Warning! Offset out of range! and/or Warning! Slope out of range!

Warning! Slope out of range!

Warning! Slope out of range!

Warning! Slope out of range! and Save

Save

Save

Save

values?

values?

values?

values? appear.

• Select between yes

yes

yes

yes or no

no

no

no.

• Press Enter

Enter

Enter

Enter to proceed.

• Select no

no

no

no to deactivate the function manual hold

manual hold

manual hold

manual hold.

• Press Enter

Enter

Enter

Enter to leave the calibration menu.

If the message ATTENTION! Slope out of range!

ATTENTION! Slope out of range!

ATTENTION! Slope out of range!

ATTENTION! Slope out of range! appears, no data can be saved.

.book Page 21 Wednesday, October 30, 2013 11:35 AM