Operation – KROHNE SMARTSENS PH 1590 EN User Manual
Page 18
4
OPERATION
18
SMARTSENS PH 1590
www.krohne.com
11/2013 - 4002545902 - MA SMARTSENS PH 1590 R02 en
• Dip the sensor into the second pH buffer solution 3 and wait till the value ist stable.
• Press Next
Next
Next
Next to proceed.
i
The measurement of pH buffer solution 2
pH buffer solution 2
pH buffer solution 2
pH buffer solution 2 starts and the values pH and temperature appear.
After approx. 10 seconds the following values appear:
pH value
temperature value
Indentified buffer value
• Press Next
Next
Next
Next to confirm the identified pH buffer solution. If neccessary change the value of the
pH buffer solution manually.
If the sensor does not recognise the pH buffer solution the message buffer not recognised
buffer not recognised
buffer not recognised
buffer not recognised
appears.
• Press Next
Next
Next
Next to proceed.
i
The value 0 for the pH buffer solution appears. Enter the value of the pH buffer solution
manually.
• Press Next
Next
Next
Next to proceed with the calibration procedure.
• Clean the sensor with demineralised water 2.
i
After a successful calibration the following values appear:
Buffer solution 1
Buffer solution 2
Offset in mV
Slope in mV/pH
• Press Next
Next
Next
Next to proceed.
• Set the calibration date with DD-MM-YYYY and calibration time with HH-MM-SS
• Press Next
Next
Next
Next to proceed.
i
The message Save values?
Save values?
Save values?
Save values? appears.
• Select yes
yes
yes
yes to save the values.
• Press Next
Next
Next
Next to proceed.
• Select no
no
no
no to deactivate the function manual hold
manual hold
manual hold
manual hold.
• Press Next
Next
Next
Next to leave the calibration menu.
A pH calibration is necessary in regular intervals or when installing a new pH sensor.
If the calibration was not successful, the message Warning! Offset out of range!
Warning! Offset out of range!
Warning! Offset out of range!
Warning! Offset out of range! appears behind
the value Offset
Offset
Offset
Offset and/or the message Warning! Slope out of range!
Warning! Slope out of range!
Warning! Slope out of range!
Warning! Slope out of range! behind the value Slope
Slope
Slope
Slope.
In case the value of the slope is not plausible, addionally the message No saving possible
No saving possible
No saving possible
No saving possible
appears.
• Press Next
Next
Next
Next to proceed.
i
The messages Warning! Offset out of range!
Warning! Offset out of range!
Warning! Offset out of range!
Warning! Offset out of range! or Warning! Slope out of range!
Warning! Slope out of range!
Warning! Slope out of range!
Warning! Slope out of range! appear.
• Press Next
Next
Next
Next to proceed.
• Set the calibration date with DD-MM-YYYY and calibration time with HH-MM-SS
• Press Next
Next
Next
Next to proceed.
i
The message Save values?
Save values?
Save values?
Save values? appears.
• Select between yes
yes
yes
yes or no
no
no
no.
• Press Next
Next
Next
Next to proceed.
• Select no
no
no
no to deactivate the function manual hold
manual hold
manual hold
manual hold.
• Press Next
Next
Next
Next to leave the calibration menu.
If the message Warning! Slope out of range!
Warning! Slope out of range!
Warning! Slope out of range!
Warning! Slope out of range! appears together with the message No saving
No saving
No saving
No saving
possible
possible
possible
possible, no data can be saved.
.book Page 18 Wednesday, October 30, 2013 11:35 AM