OPERATION

4

19

SMARTSENS PH 1590

www.krohne.com

11/2013 - 4002545902 - MA SMARTSENS PH 1590 R02 en

• Press Next

Next

Next

Next to proceed.

• Select no

no

no

no to deactivate the function manual hold

manual hold

manual hold

manual hold.

• Press Next

Next

Next

Next to leave the calibration menu.

In a fully functional sensor, the optimal slope is -59 mV for each pH unit and the optimal zero
point is 0 mV at pH 7. The slope should at least have a value between -50...-65 mV per pH unit.
Re-calibrate the sensor if the slope does not approximate those limits. The pH sensor ages, the
slope gets flatter and the zero error increases.

CAUTION!

Never touch or scratch the pH sensitive glass tip of the sensor.

Make sure that the sensor tip is clean and dust-free. If necessary, clean the tip as described
on page 24.

CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and
deliver erratic readings!
If moisture has entered the connector dry it with air (e.g. hair blower).