beautypg.com

Installation, 6 installing the sensor – KROHNE SMARTSENS PH 1590 EN User Manual

Page 16

background image

3

INSTALLATION

16

SMARTSENS PH 1590

www.krohne.com

11/2013 - 4002545902 - MA SMARTSENS PH 1590 R02 en

3.6 Installing the sensor

3.6.1 General installation instructions

The membrane glass must always have full contact with the measuring medium.

The mounting position of the sensor should not deviate more than 75° from vertical position
(sensor tip pointing downwards). Doing otherwise might cause internal air bubbles to float into
the membrane glass. This would interrupt the electrical contact between the inner buffer
solution and the glass surface.

Figure 3-6: Installation requirements

1 Measuring medium
2 Maximum deviation of 75° from vertical position

.book Page 16 Wednesday, October 30, 2013 11:35 AM