Device description, 5 protection types – KROHNE OPTIBAR PM 5060 C Ex ia EN User Manual
Page 8
2
DEVICE DESCRIPTION
8
OPTIBAR 5060
www.krohne.com
04/2014 - 4003591801 - AD OPTIBAR 5060 Ex ia R01 en
2.5 Protection types
The pressure transmitter is designed with protection type intrinsic safety, protection level "ia"
according to EN 60079-11 or IEC 60079-11.
The marking according to ATEX is:
II 1G, 1/2G, 2G Ex ia IIC T6
II 1G, 1/2G, 2G Ex ia IIC T6
II 1G, 1/2G, 2G Ex ia IIC T6
II 1G, 1/2G, 2G Ex ia IIC T6…T1 Ga, Ga/Gb, Gb
T1 Ga, Ga/Gb, Gb
T1 Ga, Ga/Gb, Gb
T1 Ga, Ga/Gb, Gb
The marking according to IECEx is:
Ex ia IIC T6...T1 Ga, Ga/Gb, Gb
Ex ia IIC T6...T1 Ga, Ga/Gb, Gb
Ex ia IIC T6...T1 Ga, Ga/Gb, Gb
Ex ia IIC T6...T1 Ga, Ga/Gb, Gb
The marking contains the following information:
The marking contains the following information:
The marking contains the following information:
The marking contains the following information:
II
IIII
II
Explosion protection, group II
1111
Equipment category 1
G
G
G
G
Gas explosion protection
Ex ia
Ex ia
Ex ia
Ex ia
Intrinsically safe, level of protection "ia"
IIC
IIC
IIC
IIC
Gas group, suitable for gas groups IIC, IIB and IIA
1/2
1/2
1/2
1/2
Equipment category 1/2
2222
Equipment category 2
Ga
Ga
Ga
Ga
EPL, suitable for zone 0
Ga/Gb
Ga/Gb
Ga/Gb
Ga/Gb
EPL, suitable for zone 0 / zone 1
Gb
Gb
Gb
Gb
EPL, suitable for zone 1
T6...T1
T6...T1
T6...T1
T6...T1
Temperature class, suitable for temperature classes T6 ... T1
.book Page 8 Thursday, June 5, 2014 3:05 PM