TECHNICAL DATA

4

23

BM 26 BASIC/ADVANCED

www.krohne.com

10/2012 - 4000346904 - QS BM 26 Basic-Adv R04 en

Measuring accuracy

Accuracy

±

10 mm / 0.4¨

(when density is constant)

≤ ±0.1% of span

±

10 mm / 0.4¨

(when density is constant)

≤ ±0.05% of span

Temperature coefficient

≤ ±0.01% of span/°C

≤ ±0.005% of span/°C

EMC immunity influence

< ±0.1% of span

< ±0.1% of span

Operating conditions

Temperature

Temperature

Temperature

Temperature

Operating temperature,

transmitter module

-40…+85°C / -40…+185°F, if there is insulation around the

measuring chamber, specify the temperature in the order. Do not put

insulation around the transmitter housing.

Operating temperature, LCD

indicator

-20…+70°C / -4…+158°F

1

Pressure

Pressure

Pressure

Pressure

Operating pressure

Atmospheric pressure

Other conditions

Other conditions

Other conditions

Other conditions

Warm-up time

5...10 minutes

30 s

Response time

1.5 s

1....60 s

2

Protection category

Transmitter housing without LCD indicator: IP 54

Transmitter housing with LCD indicator: IP 66

Installation conditions

Notes

The analog transmitter is calibrated at the factory and attached to the

measuring chamber before delivery

Dimensions and weight

Refer to the "Technical data: Dimensions and weights" section in the

Handbook

Materials

Housing

Polyester-coated aluminium

Reed-chain tube

Stainless steel

Clamp

Stainless steel

Electrical connections

Power supply

Power supply

Power supply

Power supply

Voltage

Non-Ex:

Non-Ex:

Non-Ex:

Non-Ex:

12…35 VDC

8…30 VDC

Ex ia, without LCD indicator:

Ex ia, without LCD indicator:

Ex ia, without LCD indicator:

Ex ia, without LCD indicator:

Refer to supplementary instructions or approval certificates

Ex ia, with LCD indicator:

Ex ia, with LCD indicator:

Ex ia, with LCD indicator:

Ex ia, with LCD indicator:

Refer to supplementary

instructions or approval

certificates

-

Module output

4...20 mA

4...20 mA / HART®

Order code

xF45xBxxxxx (without

LCD indicator)

xF45xExxxxx

(with LCD indicator)

xF45xWxxxxx

(without LCD indicator)

xF45xGxxxxx

(with LCD indicator)