beautypg.com

Concentration set up, 5 concentration offset – KROHNE OPTIMASS Sensors Concentration Measurement EN User Manual

Page 6

background image

2

CONCENTRATION SET UP

6

OPTIMASS

www.krohne.com

01/2014 - 4000733002 - AD OPTIMASS Concentration R04

2.5 Concentration offset

In both concentration modes, the MFC 300 and MFC 400 allows the user to set a concentration
offset. An offset may be needed where: a) there is a difference between the value measured by
the massflow meter and measurements taken in a laboratory, b) the process data does not fit
well with the converter's algorithms, or c) where there are differences between the standard
Brix values and the customer’s concentration. A good example is where pure water is not the
carrier fluid, allowing for the use of additives, fruit juices etc.The offset can be up to ±100% and
up to three decimal places, for fine corrections.

Setting the offset

Setting the offset

Setting the offset

Setting the offset

To set the concentration offset, use menu option: C2.2.2 Conc. Offset or C2.3.2 Conc. Offset and
enter the required offset value. For example, if the meter’s concentration value is 1% too low,
enter the value of +001.000 %.This will add 1% to the calculated concentration value.

AD_Concentration_en_140124_4000733002_R04.book Page 6 Tuesday, January 28, 2014 9:28 AM