CONCENTRATION SET UP

2

5

OPTIMASS

www.krohne.com

01/2014 - 4000733002 - AD OPTIMASS Concentration R04

Concentration menu

2.4 Concentration modes

The MFC 300 and the MFC 400 converters have the ability to run two concentration modes at the
same time, eg. °Brix and %Volume. It is also possible to view the percentage of Product A and the
percentage of Product B in General Concentration mode. To set the concentration mode, use
menus C2.2 Concentration 1 and C2.3 Concentration 2.

For applications where multiple products are used, the converter can be set up with two sets of
general concentration parameters. Use menus C2.4 Conc. Data 1 and C2.5 Conc. Data 2. To
select the required concentration parameter, go to menu C2.1 > Conc.Data Sel. and use the
and keys to select between Conc. Data 1 and Conc. Data 2

No

Function

Settings / description

C.2.1

Conc. Data Sel.

Define which set of general concentration parameters are active.

Select: Conc Data 1 / Conc Data 2

C2.2

Concentration 1

Defines the function of the concentration measurement. In the

following description, x=2 for Concentration 1 settings, and x=3 for

Concentration 2 settings

C2.3

Concentration 2

C2.x.1

Conc Function

Sets the concentration measurement required. Select: Off / Brix /

%Mass / %Volume / Baume 144 / Baume 145 / %NaOH / Plato / API /

%Alcohol by mass / %Alcohol by volume.

1

C2.x.2

Conc Offset

Sets an additional offset for the concentration measurement. Can be

used to correct for differences between the measured Range: -

10…10%

C2.x.3

Conc Product

Defines which general concentration product is display Select: % of

Product A / % of Product B

C2.4

Conc. Data 1

Defines the concentration coefficients used for concentration

parameter sets 1 & 2.In the following description: y=4 for Conc. Data

1 settings / y=5 for Conc. Data 2 settings.

C2.5

Conc. Data 2

C2.y.1

CCF01

Defines the use of linear or non-linear coefficients to measure

concentration. Select: Linear / Non Linear

C2.y.2

CCF02

Density of product A in g/cm

3

C2.y.3

CCF03

Temperature coefficient for Product A

C2.y.4

CCF04

Temperature coefficient squared for Product A

C2.y.5

CCF05

Defines Product B type. Select: Pure Water / Town Water / Other

C2.y.6

CCF06

Density of Product B in g/cm

3

(if CCF05 = Other)

C2.y.7

CCF07

Temperature coefficient squared for Product B (if CCF05 = Other)

C2.y.8

CCF08

Defines non-linear equation if CCF01 is set to Non Linear

C2.y.9

CCF09

Defines non-linear equation if CCF01 is set to Non Linear. Consult the

manufacturer for settings.

C2.y.10

CCF10

C2.y.11

CCF11

C2.y.12

CCF12