Concentration measurement – KROHNE OPTIMASS Sensors Concentration Measurement EN User Manual
Page 12
3
CONCENTRATION MEASUREMENT
12
OPTIMASS
www.krohne.com
01/2014 - 4000733002 - AD OPTIMASS Concentration R04
Converter settings where Product B is water
Converter settings where Product B is water
Converter settings where Product B is water
Converter settings where Product B is water
Converter settings where Product B is not water
Converter settings where Product B is not water
Converter settings where Product B is not water
Converter settings where Product B is not water
To produce the required concentration coefficiants, enter the density of the product at a number
of temperatures points. The minimum number of points required is two and the maximum is five.
For more information regarding generating concentration coefficients, refer to the Practical
considerations chapter.
No
Function
Setting
Comments
C2.x.1
Conc Function
% Mass or % Volume
C2.x.2
Conc Offset
0
If the concentration
measurement produces a stable
error, use this menu option to
enter an offset that will correct
the error.
C2.x.3
Conc Product
% of Product A or % of Product B
Select required product to be
measured
C2.y.1
CCF01
Linear
C2.y.2
CCF02
Density Product A (Solid) in
g/cm
3
C2.y.3
CCF03
0
No change with temperature
C2.y.4
CCF04
0
C2.y.5
CCF05
Pure Water or Town Water
No
Function
Setting
Comments
C2.x.1
Conc Function
% Mass or % Volume
C2.x.2
Conc Offset
0
If the concentration
measurement produces a stable
error, use this menu option to
enter an offset that will correct
the error.
C2.x.3
Conc Product
% of Product A or % of Product B
Select required product to be
measured
C2.y.1
CCF01
Linear
C2.y.2
CCF02
Density Product A (Solid) in g
/cm
3
C2.y.3
CCF03
0
No change with temperature
C2.y.4
CCF04
0
C2.y.5
CCF05
Other
C2.y.6
CCF06
Density Product B in g /cm
3
C2.y.7
CCF07
Product B Temperature term
C2.y.8
CCF08
Product B Temperature squared
term