3.2 Functional principle

The flowmeter operates in accordance with the float measuring principle.

A metal cone is installed in the measuring unit H250, in which a suitably shaped float can move
freely up and down.

The flowmeter is inserted into a vertical pipeline and the medium flows through it from bottom to
top.

The guided float adjusts itself so that the buoyancy force A acting on it, the form drag W and
weight G are in equilibrium (G = A + W).

An annular gap results which width depends on the current flow rate.

The height of the float in the measuring unit, which depends on the float, is transmitted by a
magnetic coupling and displayed on a scale.

Strong deflecting magnetic fields can lead to deviations in the measured value.