beautypg.com

Safety-related characteristics – KROHNE H250 M40 Safety V1 EN User Manual

Page 17

background image

SAFETY-RELATED CHARACTERISTICS

7

17

H250 M40

www.krohne.com

02/2012 - 4000904201 MA H250 M40 SIL R01 en

H250/M40/ESK-SE

H250/M40/ESK-SE

H250/M40/ESK-SE

H250/M40/ESK-SE with 4..20 mA current output and screwed

screwed

screwed

screwed process connections

H250/M40/HT/ESK-SE

H250/M40/HT/ESK-SE

H250/M40/HT/ESK-SE

H250/M40/HT/ESK-SE with 4..20 mA current output and screwed

screwed

screwed

screwed process connections

λ

SD

λ

SU

λ

DD

λ

DU

SFF 2 DC

D

MTBF

SIL AC 3

Stress profile 2
(low stress)

0 FIT

233 FIT

317 FIT

137 FIT

80%

70%

148 years

SIL1

Stress profile 4
(high stress)

0 FIT

261 FIT

331 FIT

172 FIT

77%

66%

133 years

SIL1

T[Proof] 4

1 year

2 years

5 years

PFD

AVG

5

6.62*E-04

1.26*E-03

3.04*E-03

λ

SD

λ

SU

λ

DD

λ

DU

SFF 2 DC

D

MTBF

SIL AC 3

Stress profile 2
(low stress)

0 FIT

243 FIT

317 FIT

157 FIT

78%

67%

142 years

SIL1

Stress profile 4
(high stress)

0 FIT

273 FIT

331 FIT

212 FIT

74%

61%

126 years

SIL1

T[Proof] 4

1 year

2 years

5 years

PFD

AVG

5

7.57*E-04

1.44*E-03

3.48*E-03

2

The number listed is for reference only. The SFF must be determined for the complete
subsystem.

3

SIL AC (architectural constraints) means that the calculated values are within the range
for hardware architectural contraints for the corresponding SIL level.

4

It is assumed that proof testing is performed with a proof test coverage of 99%.

5

The PFD

AVG

was calculated for profile 2 using the Markov modelling. The results must

be considered in combination with PFD

AVG

values of other devices of the Safety

Instrumented Function (SIF) in order to determine suitability for a specific Safety
Integrity Level (SIL)
For SIL1 applications, the PFD

AVG

value needs to be < 10

-1

.

For SIL2 applications, the PFD

AVG

value needs to be < 10

-2

.

MA_H250_M40_SIL2_R01_en_904201_PRT.book Page 17 Thursday, March 1, 2012 10:08 AM