MODBUS PROTOCOL

6

11

IFC 050

www.krohne.com

09/2012 - 4002202601 - AD Modbus IFC 050 R01 en

6.6.2 Counter controls

6.6.3 Start calibration functions

Modbus protocol
address

Description

Settings

Converter Fct. No.

3000

Start / Stop

Counter 1

Write

1

start counter

C3.1.5 / C3.1.6

Write

0

stop counter

Read

1

counter is running

Read

0

counter is stopped

3001

Start / Stop

Counter 2

Write

1

start counter

C3.2.5 / C3.2.6

Write

0

stop counter

Read

1

counter is running

Read

0

counter is stopped

3003

Reset Counter 1

Write

1

reset counter

C3.1.3

Write

0

-

Read

0

-

3004

Reset Counter 2

Write

1

reset counter

C3.2.3

Write

0

-

Read

0

-

Modbus protocol
address

Description

Settings

Converter Fct. No.

2000

Zero Calibration

Write

1

start function

C1.1.1

Write

0

-

Read

1

calibration running

Read

0

calibration finished

2001

Electrode Factor

Calibration

Write

1

start function

C1.1.8

Write

0

-

Read

1

calibration running

Read

0

calibration finished