beautypg.com

Electrical connections – KROHNE WATERFLUX 3070 Quickstart EN User Manual

Page 23

background image

ELECTRICAL CONNECTIONS

3

23

WATERFLUX 3070

www.krohne.com

06/2013 - 4002400602 - QS WATERFLUX 3070 R05 en

• Prepare appropriate cable lengths as shown.
• Connect the wires as shown in the following table.

Figure 3-4: Cable connection at converter side, standard cable

1 Connect drain wires under screw
2 Connect shielding under clamp

Wire color

Terminal

Function

Brown

1

Reference electrode

White

2

Standard electrode signal

Violet

3

Standard electrode signal

Blue

7

Field current

Green

8

Field current

Yellow

9

No function

Drain wires

Screws

Shielding