Technical data – KROHNE WATERFLUX 3000 EN User Manual
Page 27
TECHNICAL DATA
6
27
WATERFLUX 3000
www.krohne.com
06/2013 - 400162663 - HB WATERFLUX 3000 R04 en
Operating conditions
Temperature
Temperature
Temperature
Temperature
Process temperature
-5...+70°C / +23...+158°F
Ambient temperature
Standard
Standard
Standard
Standard (with aluminium converter housing):
-40…+65°C / -40…+149°F
Protect electronics against self-heating at ambient temperatures above
55°C / 130°F
Option
Option
Option
Option (with stainless steel converter housing):
-40…+55°C / -40…+130°F
Storage temperature
-50…+70°C / -58…+158°F
Measurement range
Measurement range
Measurement range
Measurement range
-12...+12 m/s / -40...+40 ft/s
Pressure
Pressure
Pressure
Pressure
Operating pressure
Up to 16 bar / 232 psi for DN25...300
Up to 10 bar / 145 psi for DN350...600
Vacuum load
0 mbar / 0 psi absolute
Pressure loss
For detailed information refer to
Pressure loss
on page 38.
Chemical properties
Chemical properties
Chemical properties
Chemical properties
Physical condition
Water: drinking water, raw water, irrigation water.
For salt water please contact the factory.
Electrical conductivity
≥ 20 μS/cm
Installation conditions
Installation
Assure that the flow sensor is always fully filled.
For detailed information refer to
Installation
on page 12.
Flow direction
Forward and reverse
Arrow on flow sensor indicates forward flow direction.
Inlet run
DN25...300:≥ 0 DN
DN350...600:≥ 3 DN
For detailed information refer to
Measurement accuracy
on page 33.
Outlet run
DN25...300:≥ 0 DN
DN350...600:≥ 1 DN
For detailed information refer to
Measurement accuracy
on page 33.
Dimensions and weights
For detailed information refer to
Dimensions and weights
on page 35.