beautypg.com

Technical data – KROHNE WATERFLUX 3000 EN User Manual

Page 26

background image

6

TECHNICAL DATA

26

WATERFLUX 3000

www.krohne.com

06/2013 - 400162663 - HB WATERFLUX 3000 R04 en

Measuring accuracy

Reference conditions

Medium: water

Temperature: +10...+30°C / +50...+86°F

Operating pressure: 1 bar / 14,5 psi

Inlet section ≥ 3 DN; outlet section ≥ 1 DN

Electrical conductivity: ≥ 300 µS / cm

Maximum measuring

error

IFC 050: 0.5% of the measured value above 0.5 m/s,

below 0.5 m/s, deviation ± 2.5 mm/s

DN25...300; IFC 100: down to 0.3% of the measured value ± 0.5 mm/s

DN25...300; IFC 300: down to 0.2% of the measured value ± 0.5 mm/s

DN350...600; IFC 100: down to 0.5% of the measured value ± 1 mm/s

DN350...600; IFC 300: down to 0.4% of the measured value ± 1 mm/s

The maximum measuring error depends on the installation conditions.

For detailed information refer to

Measurement accuracy

on page 33.

Repeatability

DN25...300 ; ±0.1% (v >0.5m/s / 1.5 ft/s)

DN350...600; ±0.2% (v >0.5m/s / 1.5 ft/s)

Calibration / Verification

Standard:

Standard:

Standard:

Standard:

2 point calibration by a direct volume comparison.

Option (for DN25...300):

Option (for DN25...300):

Option (for DN25...300):

Option (for DN25...300):

Verification to Measurement Instrument Directive (MID), Annex MI-001.

Standard: Verification at Ratio (Q3/Q1) = 80, Q3 ≥ 2 m/s

Optional: Verification at Ratio (Q3/Q1) > 80

Only in combination with the IFC 300 signal converter

MID Annex MI-001

(Directive 2004/22/EC)

EC-Type examination certificate to MID Annex MI-001

EC-Type examination certificate to MID Annex MI-001

EC-Type examination certificate to MID Annex MI-001

EC-Type examination certificate to MID Annex MI-001

Only in combination with the IFC 300 signal converter

Diameter range: DN25...300

Minimum straight inlet flow: 0 DN

Minimum straight outlet flow: 0 DN

Forward and reverse (bi-directional) flow

Orientation: any

Ratio (Q3/Q1) upto 400

Liquid temperature range: +0.1°C / 50°C

Maximum operating pressure: ≤ DN200: 16 bar, ≥ DN250: 10 bar

For detailed information refer to

Legal metrology