2

INSTALLATION

8

TIDALFLUX 2300 F

www.krohne.com

06/2013 - 4002489402 - QS TIDALFLUX 2300 R02 en

2.8.5 Open discharge

2.8.6 Cleaning of flow sensor

The flow sensor is highly resistant against dirt and the measurement will rarely be influenced by
anything. However, it is advisable to create a possiblity for cleaning just in front or behind the
sensor.

2.8.7 Flange deviation

Figure 2-10: Open discharge

1 ≥ 5 DN
2 Make sure that the water level stays below the pipe outlet.

Figure 2-11: Option for cleaning of flow sensor

1 Opening for cleaning

CAUTION!
Max. permissible deviation of pipe flange faces:
L

max

- L

min

≤ 0.5 mm / 0.02"

Figure 2-12: Flange deviation

1 L

max

2 L

min