Electrical connections – KROHNE TIDALFLUX 2300 F Quickstart EN User Manual
Page 18
3
ELECTRICAL CONNECTIONS
18
TIDALFLUX 2300 F
www.krohne.com
06/2013 - 4002489402 - QS TIDALFLUX 2300 R02 en
3.9 Preparing field current cable C, connection to measuring sensor
• The field current cable is not part of the scope of delivery.
• The shield is connected in the terminal compartment of the converter directly via the shield
and a clip.
• The shield is connected in the sensor via the special cable gland.
• Bending radius: ≥ 50 mm / 2"
Required materials
• Shielded 2-wire insulated copper cable
• Insulating tube, size according to the cable being used
• Heat-shrinkable tubing
• DIN 46 228 wire end ferrules: size according to the cable being used
1 Strip the conductor to dimension a.
2 Trim the outer shield to dimension b and pull it over the outer sheath.
3 Crimp wire end ferrules onto both conductors.
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
Figure 3-7: Preparation of field current cable C
a = 125 mm / 5"
b = 10 mm / 0.4"