Installation, 3 temperatures – KROHNE OPTIFLUX 6000 Quickstart EN User Manual
Page 15
INSTALLATION
2
15
OPTIFLUX 6000
www.krohne.com
07/2013 - 4002945701 - QS OPTIFLUX 6000 R05 en
2.9.3 Temperatures
Ambient temperature
Maximum process temperature
CAUTION!
Protect the device from direct sunlight.
°
C
°
F
min.
max.
min.
max.
Separate flow sensor
Compact version with;
IFC 050 , IFC100 & IFC 300
-40
65
-40
149
Type of connection
Separate flow sensor Compact + IFC 050
& IFC100
Compact + IFC 300
°
C
°
F
°
C
°
F
°
C
°
F
Aseptic weld on for pipes to
DIN 11850
140
284
120
1
248
2
140
284
Aseptic weld on for pipes to
ISO 2037
140
284
120
1
248
2
140
284
Dairy screw to DIN 11851
3
140
284
120
1
248
2
140
284
Screwed to SMS 1145
3
140
284
120
1
248
2
140
284
Flanges to DIN 11864-2A
140
284
120
1
248
2
140
284
Clamp joint to ISO 2852
120
248
120
248
120
248
Clamp joint to DIN 32676
140
284
120
1
248
2
140
284
Clamp joint to Tri Clamp
120
248
120
248
120
248
1 140°C if ambient temperature ≤ 40°C
2 284°F if ambient temperature ≤ 104°F
3 Without 3A mark
- BATCHFLUX 5500 C Quickstart EN (20 pages)
- IFC 050 Converter Quickstart EN (28 pages)
- IFC 100 Converter Quickstart EN (32 pages)
- IFC 300 Converter Quickstart EN (68 pages)
- OPTIFLUX 1000 Quickstart EN (20 pages)
- OPTIFLUX 2000 Quickstart EN (24 pages)
- OPTIFLUX 4000 Quickstart EN (24 pages)
- OPTIFLUX 4040C Quickstart EN (16 pages)
- OPTIFLUX 5000 Flange Quickstart EN (20 pages)
- OPTIFLUX 5000 Sandwich Quickstart EN (20 pages)
- OPTIFLUX 7300 Quickstart EN (24 pages)
- OPTIPROBE Quickstart EN (16 pages)
- TIDALFLUX 2300 F EN (44 pages)
- TIDALFLUX 2300 F Quickstart EN (24 pages)
- WATERFLUX 3000 EN (40 pages)
- WATERFLUX 3000 Quickstart EN (24 pages)
- WATERFLUX 3070 EN (80 pages)
- WATERFLUX 3070 Quickstart EN (32 pages)
- USB ADAPTER PLUS EMF EN (16 pages)
- IFC 050 Converter Modbus EN (20 pages)
- IFC 100 Converter FOUNDATION FIELDBUS EN (64 pages)
- IFC 100 Converter Modbus EN (20 pages)
- IFC 300 Converter FOUNDATION FIELDBUS EN (60 pages)
- IFC 300 Converter HART 0102 EN (20 pages)
- IFC 300 Converter HART 0201 EN (23 pages)
- IFC 300 Converter Modbus EN (24 pages)
- IFC 300 Converter PROFIBUS PA DP EN (40 pages)
- OPTIFLUX 2000-4000 IECEx EN (16 pages)
- OPTIFLUX 2000-4000-5000-6000-7300-IFC 300 Ex EN (40 pages)
- OPTIFLUX 2000-4000-5000-6000 -IFC 100 Ex EN (24 pages)
- OPTIFLUX 4040 C Ex EN (20 pages)
- OPTIFLUX x300 Ex Zone2 EN (1 page)
- H250 M9 ES EN (36 pages)
- VA 40-VA 45 EN (36 pages)
- H250 M10 ATEX II2G Ex d EN (16 pages)
- H250 M10 ATEX II3D Ex t EN (16 pages)
- H250 M40 ATEX II2D Ex t-II2G Ex d EN (20 pages)
- H250 M40 ATEX II2G Ex i EN (20 pages)
- H250 M40 ATEX II3G Ex nA EN (20 pages)
- H250 M40 Ex II2G Reed EN (2 pages)
- H250 M9 ATEX II2G Ex i EN (16 pages)
- H250 M9S ATEX II3D Ex t-II3G Ex nA EN (20 pages)
- M8E Converter HART 0101 EN (13 pages)
- DK 32-DK 34 ATEX II2G Ex i EN (16 pages)