2.8.9 Mounting requirements for self-draining

Minimum slope

Figure 2-15: Vacuum

1 ≥ 5 m

INFORMATION!
Applicable for 3A marked installations: install flow sensor in vertical pipelines or in pipelines
with a minimum slope as indicated!

Figure 2-16: Installation note for 3A marked installations

1 Minimum slope

Nominal

diameter

DIN

11850

ISO

2037

DIN

11864 2A

ISO

2852

DIN

32676

Tri

Clamp

2.5...6

10°

10°

-

-

-

-

10

-

-

-

-

15

10°

10°

-

-

-

-

25

10°

10°

10°

40...50

65...80

10°

10°

10°

100

125...150

10°

10°

-

-

1