ASCII SO Opens the dark shutter which allows monochromatic light to enter the

sample chamber. This allows 100% transmittance (zero absorbance) to
be calibrated. The shutter must be open for normal measurements.

ASCII Gnnn Commands the instrument to go to wavelength nnm. For example:

G540 will set the wavelength to 540nm.

ASCII Fxxxx.x Sets the concentration factor to xxxx.x. For example: F1000 will set

the factor to 1000.

Note is an ASCII carriage return character. The last three commands provide an output

which can be readily incorporated into most spreadsheet software packages.

6.1

RS232 Output

The bi-directional RS232 interface is available on the rear panel 25 way D type connector. The
connections are as follows:

TDX 2

– INPUT TO 6305

RXD 3

– OUTPUT FROM 6305

RTS 4

– LINKED TO CTS

CTS 5

– LINKED TO RTS

DSR 6

– OUTPUT FROM 6305

DCD 8

– OUTPUT FROM 6305

DTR 20

– INPUT TO 6305 (must be active)

GND 7

Suggested interconnections are detailed below:

6305

IBM PC XT (25 way “D”)

TDX 2

2 TXD (From PC)

RXD 3

3 RXD (To PC)

RTS 4

4 RTS (From PC)

CTS 5

5 CTS (To PC)

DSR 6

6 DSR (To PC)

DCD 8

8 DCD (To PC)

DTR 20

20 DTR (From PC)

GND 7

7 GND