Jenway 6305 User Manual
Page 21
21
calibrated.
ASCII SO
sample chamber. This allows 100% transmittance (zero absorbance) to
be calibrated. The shutter must be open for normal measurements.
ASCII Gnnn
G540
ASCII Fxxxx.x
the factor to 1000.
Note
which can be readily incorporated into most spreadsheet software packages.
6.1
RS232 Output
The bi-directional RS232 interface is available on the rear panel 25 way D type connector. The
connections are as follows:
TDX 2
– INPUT TO 6305
RXD 3
– OUTPUT FROM 6305
RTS 4
– LINKED TO CTS
CTS 5
– LINKED TO RTS
DSR 6
– OUTPUT FROM 6305
DCD 8
– OUTPUT FROM 6305
DTR 20
– INPUT TO 6305 (must be active)
GND 7
Suggested interconnections are detailed below:
6305
IBM PC XT (25 way “D”)
TDX 2
2 TXD (From PC)
RXD 3
3 RXD (To PC)
RTS 4
4 RTS (From PC)
CTS 5
5 CTS (To PC)
DSR 6
6 DSR (To PC)
DCD 8
8 DCD (To PC)
DTR 20
20 DTR (From PC)
GND 7
7 GND