Caution – HT instruments HT401 User Manual
Page 13
![background image](/manuals/367495/13/background.png)
HT401
EN - 12
4.4.4. DC Current measurement
CAUTION
The maximum input DC current is 10A. Do not measure currents exceeding
the limits given in this manual. Exceeding these limits could result in
electrical shocks to the user and damage to the instrument
Fig. 5: DC current measurement
1. Cut off power supply from the circuit to be measured.
2. Select the
or positions
3. Insert the red cable and the black cable into the input terminals mA or A and COM
4. Press the MODE key to activate “DC” or “AC+ DC” modes (see § 4.3.2)
5. Press the RANGE key to activate manual range selection (see § 4.2.1) or use the
Autorange feature. If the current value is unknown select the maximum range
6. Connect the red lead and the black lead in series to the circuit whose current you want
to measure, respecting polarity and the current direction indicated in Fig. 5
7. Supply the circuit to be measured. The display shows the value of current.
8. The message “O.L” indicates that the current being measured exceeds the maximum
value measurable by the instrument
9. When symbol “-” appears on the instrument’s display, it means that current has the
opposite direction with respect to the connection in Fig. 5
10. For Maximum and Minimum value measurement and HOLD function measurement,
please refer to § 4.2