NOTE 2: If the sensor fails the stability criteria, a “Stability Fault” will be declared and a “Fault
Detected
” message will be displayed alternately on the sensor with the current reading. The HART
Sensor Status will change to reflect a Sensor Fault.

NOTE 3: If the sensor fails the clearing time criteria, a “Clearing Fault” will be declared and a “Fault
Detected
” message will be displayed alternately on the sensor with the current reading. The HART
Sensor Status will change to reflect a Clearing Fault.

4.3.2.2 Loop Test Calibrate


The Calibration Screen contains a Loop Test Calibration section. This section displays the loop current
reading and allows the user to perform D/A trim and Loop Test. D/A trim is used to calibrate the 4mA and
20mA set points in the Digital/Analog converter. Loop test allows the user to set the mA output to any level
within range; this can be a good troubleshooting tool for line degradation. A DVM capable of reading
milliamps and a 100~250 Ohm resistor are required to perform D/A trim and/or Loop test.

HRT BRIDGE

SENSOR

+24V

mA/HART

mA/HART

GND

+24V

CFG

HART

HART

+ 24V

100~250 Ohm Resistor

DVM +mA

GND (- 24V)

DVM GND

HART

Connector

HART

Header

HRT Bridge PCA

Figure 8 DVM Connection

D/A Trim

1) Select “D/A trim” from the Loop test Calibrate section of the screen.
2) The procedure will prompt to connect a meter: (Refer to Figure 8.)

a) Disconnect the wire from the HART Connectors mA/HART position.
b) Connect one end of a 100~250Ohm resistor to the mA/HART position of the HART Connector.
c) Connect the positive lead of a DVM set to measure milliamps to the other end of the resistor.
d) Connect the negative lead (-) of the DVM to the GND position of the connector.

3) The procedure will set the output to 4ma, and prompt for the DVM Reading. Input the reading and select

“OK”.

4) The HRT Bridge will readjust the output for 4mA, and inquire if the reading is 4mA on the DVM (select

“YES” or “NO”).

5) If the reading on the DVM is 4mA, select “YES” and “OK” and the HRT Bridge will continue on to

20mA. If the reading is not 4mA, select “NO” and “OK” and the procedure will re-run the 4mA
calibration (Step 3).

6) The procedure will set the output to 20mA, and prompt for the DVM Reading. Input the reading and

select “OK”.

7) The HRT Bridge will readjust the output for 20mA, and inquire if the reading is 20mA on the DVM

(select “YES” or “NO”).

8) If the reading on the DVM is 20mA, Select “YES” and “OK”. If the reading is not 20mA, select “NO”

and “OK” and the procedure will re-run the 20mA calibration (Step 6).