Figure 7 fp configuration setup – Detcon 700 HRT Bridge User Manual
Page 13
HRT Bridge
700 Sensor HRT Bridge IM
Rev. 0.2
Page 9 of 18
Figure 6 FP Configuration Setup
This screen will vary depending on the type of Model 700 sensor attached to the HRT Bridge. The Range Set
up will display the Model Type of the sensor, followed by the PV LRV (Primary Variable Lower Range
Value) and the PV URV (Primary Variable Upper Range Value), and the Conc Units (Concentration Units),
the display may also show the Sensor Range. The Sensor Setup portion of the screen will display sensor
specific parameters:
DM
• Sensor Range
• Cell Bias
• Gain Code
• Raw Counts
• Sensor Life
FP
• Gas Factor
• Cal Factor
• Bridge Current
• Bridge Voltage
• Sensor Life
IR
• Gas Factor
• Active Counts
• Reference Counts
• Range Multiplier
• Sensor Life
PI
• Sensor Range
• Gain Code
• Raw Counts
• Zero Offset
• Sensor Life