Notice for User
Please use the unit in ranges of applicable temperature, humidity and pressure
that are appropriate for the specification of the product. Using the unit
beyond this range may cause a malfunction/glitch of the unit.
Gas concentration measurement values by the sensor or the unit can vary with
respect to the environment at the site (temperature, pressure and humidity).
Therefore, the calibration of the unit should be performed at the same or similar
environment as that of the unit use (temperature, pressure and humidity).
If temperature changes sharply during use of the unit (for instance, using the unit
where there are varying temperatures between indoor and outdoor), the value of
the measured gas concentration can change suddenly. Please allow time for
the gas concentration value to stabilize.
Severe vibration or shock to the unit may cause a sudden change in value of the
measured gas concentration. Please allow time for the gas concentration value to
stabilize. Excessive shock to the unit can lead to problems with the sensor or the
unit.
Parts List
Part Number
Description
Part Number
Description
96P-240000-100
Model SG1-H2S,
Range 0-100ppm
96P-840000-01K
Model SG1-H2,
Range 0-1,000ppm
96P-240000-200
Model SG1-H2S,
Range 0-200ppm
96P-340000-030
Model SG1-O2,
Range 0-30%
96P-440000-500
Model SG1-CO,
Range 0-500ppm
37P-240000-000
H2S Plug-In Sensor
96P-440000-01K
Model SG1-CO,
Range 0-1,000ppm
37P-440000-000
CO Plug-In Sensor
96P-550000-020
Model SG1-SO2,
Range 0-20ppm
37P-550000-000
SO2 Plug-In Sensor
96P-550000-100
Model SG1-SO2,
Range 0-100ppm
37P-840000-000
H2 Plug-In Sensor
96P-840000-500
Model SG1-H2,
Range 0-500ppm
37P-340000-000
O2 Plug-In Sensor