When the 1st alarm occurs, the operator can acknowledge it by pressing the Power key. Only

the alarm sound stops. The LED will stay on until the reading drops below the alarm value.
When the 2nd alarm occurs, the operator and workers should promptly leave the work site.

The operator can acknowledge the alarm by pressing the Power key in a place where the

concentration value of the gas is normal. (The alarm alerts continuously when it does not

return to a normal state even if the Power key is pressed.)
When STEL / TWA alarm occurs, it is indicated with the value of the measured

concentration and activates the alarm with the same sound and vibration as that of the

2nd alarm.

When STEL / TWA alarm occurs, their respective icons can be deleted only by Power Off.

Primary battery alarm sounds repeatedly at 5 minute intervals only when one bar of the battery

icon remains.

Secondary battery alarm starts right before the end of power, and the power source finishes

after 10 seconds from the outbreak of alarm.

In the event of a test or calibration failure, the

icon is displayed with an alarm sound.

Alarm Set Point

Gas

O

2

CO

H

2

S

H

2

SO

2

1

st

19%

30 ppm

10 ppm

100 ppm

2 ppm

2

nd

23%

60 ppm

20 ppm

500 ppm

4 ppm

TWA

N/A

30 ppm

10 ppm

N/A

2 ppm

STEL

N/A

200 ppm

15 ppm

N/A

5 ppm

Alarm Set

To enter Alarm Set Mode, press the Arrow key for 3 seconds while the unit is powered off.
SET will be displayed and the

icon will blink. Press the Power key to enter the alarm 1

value. Use the Arrow key to change the value and the Power key to move to the next

digit. Pressing the Power key while on the last digit will return to the alarm 1 SET screen.

Hold 3 secs

10