3.3 S

PECIFICATIONS

Method of Detection

Solid state metal oxide diffusion/adsorption

Electrical Classification

CSA-NRTL (US OSHA) approved Class I; Groups B, C, D; Div. 1.

Response Time

T50 < 30 seconds, T80 < 60 seconds

Clearing Time

T80 < 30 seconds

Repeatability

± 10% of reading or ± 2 ppm (greater of)

Range

0-20 ppm, 0-50 ppm, or 0-100 ppm

Operating Temperature

-40° to +175° F (-40°C to +75°C)

Accuracy

± 10% of reading or ± 2 ppm (greater of)

Sensor Warranty

10 year conditional

Power Consumption

Normal operation = 68 mA (<1.7 watt); Full alarm = 85 mA (2 watts)

Output

Linear 4-20 mA DC

Input Voltage

12-28 VDC

3.4 O

PERATING

S

OFTWARE

Operating software is menu listed with operator interface via the two magnetic program switches located under the

face plate. The two switches are referred to as “PGM 1” and “PGM 2”. The menu list consists of 3 items which

include sub-menus as indicated below. (Note: see the last page of this manual for a complete software f low chart.)

01.

Normal Operation

a) Current Status

02.

Calibration Mode

a) Span

03.

Program Menu

a) View Program Status

b) Set Calibration Level

c) Set Heater Level

d) Set Range

e) Utility Menu

3.4.1 Normal Operation

In normal operation, the display ref lects the current status of the sensor and gas concentration and appears as:

“0 PPM H2S”

. The mA current output corresponds to the monitoring level and range of detection = 4-20 mA.

3.4.2 Calibration Mode

Detcon Model TP-524C Hydrogen Sulfide Sensor