Setup reset set clock units flip display – DeFelsko PosiTector PC v.1.1 User Manual
Page 14
13
Setup Menu
Reset restores factory settings and returns the gage to a known, out-of-
the-box condition. It is handy when you want to "start all over".
The following occurs:
- All batches are closed and stored measurements are erased.
- Calibration adjustments are cleared and the instrument is returned to
the gage's factory calibration settings as CAL 1.
- Menu settings are returned to the following:
Memory = OFF
Statistics = OFF
USB Drive = ON
A more thorough Reset can be performed by holding the (+) button
when the gage is powered down until the Reset symbol appears. It
performs the same function as a menu Reset with the addition of Units
= microns, Language = English, and Flip Display = Normal.
A probe Zero (pg. 6) must be performed after any Reset.
This option causes the base unit display to read upside down. Ideal for
viewing while the base unit is attached to a waist belt so that the display
is conveniently pointed toward the operator.
This menu button converts the display and all stored readings from
microns to mils or vice versa.
All batches are date-stamped when created and all measurements are
time-stamped (24 hour format) when stored into these batches. It is
therefore important to keep both the date and time current using this
menu option.