echnical Dat

echnical Dat

a

a

Conforms to: ISO 8502-4

Range

Accuracy

Resolution

Surface Temperature

-40° to 80° C ±0.5° C 0.1° C

80° to 190° C ±1.5° C 0.1° C

-40° to 175° F ±1° F 0.1° F

175° to 375° F ±3° F 0.1° F

Air Temperature

-40° to 80° C ±0.5° C 0.1° C

-40° to 175° F ±1° F 0.1° F

Humidity

0 to 100% ±3 % 0.1 %

Operating range

-40° C to +80° C (-40° F to +175° F)

Physical Specifications:

Body: 147 x 61 x 25mm (5.8" x 2.4" x 1") Probe: 220 x 20mm (8.7" x 0.8")

Weight: 170g (6oz) without batteries Battery Life: 50 hours continuous

Returning for Service

Returning for Service

Before returning the instrument for service…

1. Install new Alkaline batteries in the proper alignment shown on the

door.

2. Examine the surface temperature probe tip for dirt or damage.

3. Perform a RESET. (pg. 10)

To return an instrument for service print out a service form on
our web site.

Why do the Surface Temperature and Air Temperature values differ

when holding the probe in the air and away from a surface?

The Ts sensor has a very quick response time. The Ta
sensor has a slower reaction time and requires additional
time to stabilize. This is most noticable when moving from
one temperature extreme to another.

Changing The Batteries

Changing The Batteries

As the batteries become weak the

symbol will appear

and begin flashing during measuring. If the batteries are
allowed to become very weak, this symbol will remain on
without flashing.

USE ONLY AA ALKALINE BATTERIES.

Nickel-

cadmium and nickel-metal hydride rechargeable batteries will
work but the low battery symbol may stay on.