2 technical data, weighing, Optional accessories – Seca 635 User Manual
Page 99
Optional accessories • 99
English
12.2 Technical data, weighing
13.OPTIONAL ACCESSORIES
Wireless transmission
• Frequency band
• Transmission power
• Standards applied
2.433 GHz - 2.480 GHz
< 10 mW
EN 300 328
EN 301 489-1
EN 301 489-17
General technical data
seca 635
Technical data, weighing
Calibration in accordance with Directive
2009/23/EC
Class III
Maximum load
• Partial weighing range 1
• Partial weighing range 2
200 kg
300 kg
Minimum load
• Partial weighing range 1
• Partial weighing range 2
2.0 kg
4.0 kg
Increments
• Partial weighing range 1
• Partial weighing range 2
100 g
200 g
Tare range
300 kg
Accuracy on initial calibration
• Partial weighing range 1, to 50 kg
• Partial weighing range 1, 50 kg to 200 kg
• Partial weighing range 2, to 100 kg
• Partial weighing range 2, 100 kg to 300 kg
±50 g
±100 g
±100 g
±200 g
seca 360° wireless devices
Article number
Length measuring rods
•
seca 274
•
seca 264
Country-specific versions
Country-specific versions
Wireless printers
•
seca 360° wireless printer 465
•
seca 360° wireless printer advanced 466
Country-specific versions
Country-specific versions
PC software
•
seca analytics 115
Application-specific license
packages
seca 360° wireless USB adapter 456
456-00-00-009