11.SERVICING

On leaving the factory, your seca scale has an accu-
racy of ±0.15 % or better. To preserve this level of
accuracy, the product must be set up with care and
serviced regularly. Depending on how frequently the
scale is used, we recommended servicing at intervals
of 3 to 5 years.

ATTENTION!

Incorrect measurements as a result of poor
servicing

– Have servicing and repairs carried out

exclusively by seca service or by an
authorized service partner.

– You can find service partners in your area at

www.seca.com or by sending an e-mail to
[email protected].

The display "Er:X:16"
appears.

The scale was caused to oscillate, the zero point
could not be determined.

- Restart the scale.

The Enter key
(send/print) is pressed
and the display "Er:X:71"
appears.

No data transmission possible, wireless module
deactivated.

- Activate wireless module (see "Activating the

wireless module (SYS)" on page 88).

The Enter key
(send/print) is pressed
and the display
"Er:X:72" appears.

No data transmission possible, no wireless group
set up.

- Set up wireless group (see "Setting up a

wireless group (Lrn)" on page 88).

Problem

Cause / Solution