Connecting the connecting cable to the platform – Seca 634 User Manual
Page 71
Before you really get started… • 71
English
Connecting the
connecting cable to the
platform
NOTE
When selecting the setup location, take
account of the length of the connecting cable
so that you can place the display housing on
the wall bracket.
1. Place the scales on a firm, level surface.
2. Connect the connecting cable to the electronics
box of the scales.
ATTENTION!
Incorrect measurement due to force shunts
If the cable of the display housing contacts the
weighing platform, the patient's weight cannot
be measured correctly.
– Route the cable in such a way that it cannot
contact the weighing platform.
See also other documents in the category Seca Scales:
- 334 (24 pages)
- 336 (115 pages)
- 804 (166 pages)
- 354 (2 pages)
- 376 (401 pages)
- 384 (136 pages)
- 745 (56 pages)
- 717A (136 pages)
- 757 (307 pages)
- 985 (239 pages)
- 955 (19 pages)
- 955 (35 pages)
- 956 (136 pages)
- 959 band_1 (317 pages)
- 700 (2 pages)
- 703 (77 pages)
- 703 (102 pages)
- 799 (162 pages)
- 786/culta/cultura (38 pages)
- 769 (76 pages)
- 704 (443 pages)
- 634 (77 pages)
- 635 (188 pages)
- 635 band_1 (369 pages)
- 761 (155 pages)
- 754 (8 pages)
- 761 (156 pages)
- 899 (188 pages)
- 878 (139 pages)
- 874 (65 pages)
- 869 (102 pages)
- 657 band_1 (357 pages)
- 657 (240 pages)
- 645 (417 pages)
- 665 (188 pages)
- 665 band_1 (377 pages)
- 677 band_1 (337 pages)
- 677 (240 pages)
- 685 (292 pages)
- 685 band_1 (385 pages)
- 675 band_1 (292 pages)
- 675 (214 pages)
- 856 (65 pages)
- 757 band_1 (307 pages)