Cleaning, What to do if – Seca 959 User Manual
Page 83
Cleaning • 83
English
8. CLEANING
Clean the seat and frame on an as needed basis using
a household cleaner or a commercial disinfectant. Fol-
low the manufacturer's instructions.
Do not use scouring or caustic cleaning products,
spirit, petrol or similar substances for cleaning under
any circumstances. Such products may damage the
high-quality finishes.
9. WHAT TO DO IF …?
Malfunction
Cause/Remedy
… no weight is displayed
during weighing?
The scale has no power supply.
- Check whether the scale is switched on
- Check whether the rechargeable battery
block is inserted
…
0.00
does not appear
before weighing?
A load was placed on the scales before being
switched on.
- Remove load
- Switch scales off and back on again
… a segment lights up
continuously or not at
all?
There is a fault at that point.
- Notify maintenance service.
…
is displayed?
The voltage of the rechargeable battery block
drops.
- Charge the rechargeable battery block as
soon as possible
…
bAtt
is displayed?
Rechargeable battery block is empty.
- Charge rechargeable battery block
…
St0p
is displayed?
The maximum load has been exceeded.
- Reduce load
…
is displayed?
The ambient temperature of the scales is too high
or too low.
- Set up scales in an ambient temperature
between +10 °C and +40 °C
- Wait approx. 15 minutes until scales have
adjusted to ambient temperature