Maintenance, Technical data – Seca 985 User Manual
Page 35
Model 985
35
GB
… the display
Er:1:11
appears?
– At least one weighing cell is not correctly connected or the connection is back to front.
Check the connections. The colours of plug and socket must match.
Switch the scale off using the Start key and restart. The scale then works normally
again. If this is not the case, inform Servicing.
… the display
Er:1:12
appears?
– The maximum weight for switching on was exceeded when the scale was switched
on, e.g. if bed and patient are being weighed. Remove the load from the scale and start
it again. The scale then works normally again. If this is not the case, inform Servicing.
… one segment is permanently illuminated or is not illuminated at all?
– The relevant segment is faulty. Inform Servicing.
8. Maintenance
Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged or the calibra-
tion counter does not match the number on the applicable calibration counter mark (cf.
page 27).
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.
9. Technical data
Dimensions
width:
540 mm
height:
990 mm
depth:
540 mm
weight
approx. 25 kg
Measuring feet
width:
328 mm
height:
150 mm
depth:
212 mm
weight:
approx. 6.2 kg
Cable length:
2.80 m
Size of figures in display
20 mm
Temperature range
+10 °C to +40 °C
Calibration
as per Directive
2009/23/EC
approved for medical use,
Class III
Medical product as per Directive
93/42/EEC
Class I with measuring function
Maximum load
250 kg