Storing the weight value (hold) – Seca 336 User Manual
Page 15

Model 336
15
GB
Storing the weight value (HOLD)
– Press the HOLD key whilst the load is
on the scale. 
The display flashes until a stable weight 
has been measured. The display is then 
frozen until the next weighing operation 
takes place.
– “HOLD” and
appears in the digital
display. The triangle with the 
exclamation mark indicates that this 
frozen display is an additional facility 
which is not subject to calibration.
– The HOLD function can be switched off
by pressing the HOLD key again.
0.0
appears in the display if the scale
has no load on it, otherwise the current 
weight. The “HOLD“ display vanishes.
Note: 
The HOLD function can also be used if the TARE function has been executed 
beforehand.
Checking that the contents of the calibration counter are correct
This seca scale is calibrated. Calibration may be performed only by authorised centres. 
To ensure this, the scale is fitted with a calibration counter which records any change to 
data relevant to calibration. If you want to check whether the scale is properly calibrated, 
proceed as follows:
– If necessary, switch off the scale.
– Keep any key depressed and start the 
scale.
– The current contents of the calibration
counter are shown on the display for a 
few seconds. 
– Compare the output contents of the
calibration counter with the number 
shown on the calibration counter sticker. 
For valid calibration, both numbers must 
match. If sticker and calibration counter do 
not match, then re-calibration is required. 
Contact your service agent or seca Cus-
tomer Service.
Both numbers must match.