UPS3000/UPS3110/UPS3210 Operation and Maintenance Manual

3.7.1

For UPS3000/UPS3210 Absolute/Gage Switch Selectable Units ONLY

If the current barometric pressure is below 14.7 PSIA, the offset is positive, see Example 1. If the current
barometric pressure is above 14.7 PSIA, the offset is negative, see Example 2. If the current barometric pressure
is 14.7 PSIA, then no offset is needed.

Example 1:

If the current barometric pressure is lower than 14.70 PSIA, subtract the current barometric pressure from 14.70.

14.70 PSI: UPS3110 reference point

-14.55 PSI: Current barometric pressure

0.15 PSI: Positive Delta Offset

Complete the following steps

NOTE: UPS3000 refer to Figure 2-1 on page 5 and Figure 2-4 on page 6. UPS3210 refer to Figure 2-3 on page 6 and
Figure 2-6 on page 8
.

1. Connect Test Standard to UPS3000/UPS3210 Input Port similar to Figure 3-1 on page 19.

2. Open the vent valve connected between Test Standard and UPS3000/UPS3210.

3. Using the

RANGE SELECT

switch (6), select the lowest pressure range on the UPS3000/UPS3210.

4. Press the

ENTER

push-button, as shown in Figure 3-2 on page 20, on the UPS3000/UPS3210. The

display reads zero. Repeat this step for the middle and high ranges.

5. Close the vent

valve connected between Test Standard and UPS3000/UPS3210. Select the lowest

pressure range of the UPS3000/UPS3210.

6. Input a pressure into the UPS3000/UPS3210 until the display reads the value of the Positive Delta

Pressure, 0.15 PSI as in this example.

7. Depress the

ENTER

push-button on the UPS3000/UPS3210. The display reads zero. Without touching the

input pressure

repeat this step for the middle and high ranges.

8. Using the

STEPPER

push-button, as shown in Figure 3-2 on page 20, place the unit into

SHUNT RESISTOR

CALIBRATION

mode. The display will be as shown in Figure 3-5 on page 23.

9. With UPS3000/UPS3210

RANGE SELECT

switch (6) switched in the high range, depress the

ZERO

push-button on the front panel. Verify that the display reads 0.00. Upon release of the push-button the
display will be as shown in Figure 3-5 on page 23.

10. Depress the

ENTER

push-button. The display will respond with

“[] [] [] [] [] []”

until the button is

released.

11. Pressing the

STEPPER

push-button bring the indicator back to its initial Data Recall display condition as

shown in Figure 3-2 on page 20. The display shows

1 XXX

(three arbitrary digits).

12. Depress the

DATA ENTER

switch, S2, located approximately in the middle of the circuit board. See Figure

2-7 for the switch location.

13. If the data is accepted, the three-digit number on the right side of the display will indicate

377

for as long

as the

DATA ENTER

switch, S2, is depressed.

14. The barometric offset is now complete and

CALIBRATE/OPERATE

switch, S3, must be returned to its

normal operating positions as shown in Table 3-1 on page 20. The Test Standard may now be
disconnected.