Assignment of Power Supply Terminals
1
1
2
3
2
3
Terminal Number
Signal Symbol
SUPPLY
L (
ϩ
)
SUPPLY
N(
Ϫ
)
GND
CAUTION
Ensure that the power supply voltage for VJCE-01A matches that for the signal conditioner to be
mounted on VJCE-01A. Supply of different power supply voltage may damage VJCE-01A and signal
conditioners.
6. CALIBRATION
Refer to the User’s Manual of respective signal conditioners for how to calibrate and the equipment required for cali-
bration.
6.1
Items to Check before Power on
●
Supply power rating is 12 to 36 V DC or 85 to 246 V AC / DC.
●
Wiring of signal cables
●
Installation, ambient temperature, humidity, dust and vibration
Please power on after checking the above items.
The VJCE-01A would be in operational status upon power on. However, 10 to 15 minutes are required to satisfy its
specifications and performance