Yokogawa PC-Based MX100 User Manual
User’s manual, Warning, Attachment procedure
User’s
Manual
Yokogawa Electric Corporation
IM MX100-74E
6th Edition
WARNING
• To prevent electric shock and instrument breakdown, do not connect the power
supply to the main module when attaching modules.
Attachment Procedure
1. Check that the power supply is not connected to the main module.
2. Align the connector on the rear panel of the module to the connector at the desired
position of the base plate and insert the connector.
When the connectors are correctly connected, the guide pin on the rear panel of the
module is inserted into the guide hole on the base plate. In addition, the module is
secured to the base plate with the latch lever locking in place at the bottom section of
the base plate.
3. Fasten the attachment screws (M3) at the top of the module.
To remove the module, loosen the attachment screw, pull down on the latch lever on the
rear panel of the module, and pull the module straight from the base plate.
*1
Base plate connector cover
*1
(accessory sold separately)
Input/Output module
Module attachment screw
Module connector
*2
Screw holes
Guide holes
MX100
Main module
Base plate
Base plate
Latch lever
(Pull up to remove the module
from the base plate.)
Guide pins
MW100
Main module
Screw holes
Guide holes
Module connector
*2
Base plate
Base plate connector cover
*1
(accessory sold separately)
Input/Output module
Module attachment screw
Base plate
Latch lever
(Pull up to remove the module
from the base plate.)
Guide pins
MX100 Main module
MW100 Main module
Can be attached by hooking the top section of the cover to the top edge of the base plate and pressing down on the lower section of the
cover
When attaching the module, be sure that the module
connector is vertical to the base plate so that the
module connector pins are not bent.
*2
Note
• Attach the main module to the right-most position on the base plate.
• The 30-CH, Medium Speed DCV/TC/DI Input Module takes up three modules worth of
space when attaching to the base plate. If attached incorrectly, damage or malfunction can
result.
• The 10-CH, Pulse Input Module is compatible with the MW100.
6th Edition: September 2013 (YK)