4 mounting the isc40, 3 mounting the sc4a – Yokogawa PR10 Retractable Assembly User Manual
Page 15
15
IM 12B6K3-01E-E
3.8.4 Mounting the ISC40
1 Take the sensor out of the box and
remove the cable tie carefully
2 Bind the separate wires of the cable
together with a piece of tape
3 Take the fitting out of the box and
remove the option(s), if necessary
4 Release the pigtail (cable gland)
completely. Do not undo the part in the
metal tube!
5 Lead the sensor cable through the tube
of the fitting, from the side where the
knurled knob has been removed
6 Hold the sensor still and turn the metal
tube onto the sensor. Don’t rotate the
cell, but rotate the tube of the fitting,
because the cable can be disconnected
from the cell, when rotating it
7 Lead the loose part of the pigtail onto
the cable and screw it onto the fixed
part
8 Remove the tape
5
FIG. 10 Mounting the ISC40
FIG. 9 Mounting the SC4A
3.8.3 Mounting the SC4A
1 Take the sensor out of the box and cut
off the cable tie
2 Bind the separate wires of the cable
together with a piece of tape.
3 Take the fitting out of the box and
remove the option(s), if necessary.
4 Release the pigtail (cable gland)
completely. Do not undo the part in the
metal tube!
5 Lead the sensor cable through the tube
of the fitting, from the side where the
knurled knob has been removed.
6 Hold the sensor still and turn the metal
tube onto the sensor. Don’t rotate the
cell, but rotate the tube of the fitting,
because the cable can be disconnected
from the cell, when rotating it
7 Lead the loose part of the pigtail onto
the cable and screw it onto the fixed
part.
8 Remove the tape.
5